Newton Scientific Fume Hood
Newton Scientific ducted fume hood protect user from being exposed to toxic fumes and hazardous particulate by pulling out contaminated air via the building’s HVAC system and out of the building. This allows contaminated air inside the laboratory to be pushed out from the building and ensure polluted air is not recirculated back into the laboratory.
Features:
- Double layered design
- Aerodynamic front fascia
- Industrial grade galvanised steel construction
- Heavy epoxy powder coated finish to withstand harsh laboratory environments
- Phenolic resin inner liner and baffle with built in slots for efficient scavenging
- Dished epoxy resin worktop with high chemical and heat resistance, black
- Purpose built light box external to airflow, preventing any corrosion and minimizing turbulence
- Electronic light ballast for flicker free usage and user eye comfort
- Counterbalanced sash assembly for years of trouble free use
- Tempered safety glass with Easygrip® hand rail
- 2 side access panels built into inner liner for future addition of utilities
- Stainless steel airfoil designed to maximize air sweep of work surface
- All fume hoods comply to exhaust requirements and face velocity test as per ASHRAE 110-95; when ordered with control systems, units also comply to BS 7258 and EN 14175
Product Specification
Model
4 Feet, Ducted
5 Feet, Ducted
6 Feet, Ducted
8 Feet, Ducted
Cat. No.
61000660
61000661
61000662
61000663
Dimensions, mm (LxWxH) Including base cabinet
1220 x 920 x 2420mm
1520 x 920 x 2420mm
1820 x 920 x 2420mm
2420 x 920 x 2420mm
Body
Electrogalvanised steel with heavy epoxy powder coated finish
Worktop
Dished epoxy resin, black with marine edge
Inner Liner & Baffle
Phenolic resin, laboratory grade
Airfoil
SS 304
Standards Complied
• ASHRAE 110-95
• BS 7258 (with Optional Controller)
• EN 14175 (with Optional Controller)
Services
• 1 Pre Plumbed Cold Water Remote
• 1 Pre plumbed Gas Remote
• 2 x Pre wired 13 Amp Switch socket outlets (BS)
• PP Cup Sink
Controls
Fan ON/OFF switc, Light ON/OFF switch
Base
Epoxy powder coated steel construction
Duct Collar
200 mm
250 mm
300 mm
300 mm x 2
Design Velocity
0.5 m/s at 500mm sash height
Extract Volume at 500 mm
Height Sash Opening
1188 CMH
1512 CMH
1836 CMH
2481 CMH
Product Specification
Top View
1. PVC Exhaust Collar
2. Electrical Junction Box
Side View
3. External Light Box
4. Side Access Panel
5. Stainless Steel 304 Airfoil
Front View
6. Removable Front Panel
7. Tempered Glass Sash
8. PP Cup Sink
9. Epoxy Resin Worktop
10. Fan ON/OFF Switch
11. Light ON/OFF Switch
12. Gas Fixture Remote
13. Water Fixture Remote
14. Gas Faucet
15. Water Faucet
16. Electrical Outlets